Name Description
[MF2 Speed] Threshold
[mm/s]
Enter speed threshold. The switching hysteresis is fixed at 100 mm/s.
Options
•
0 ... 15,000 [mm/s]
Factory setting
•
5,000 [mm/s]
[MF2 Speed] Mode Select direction for detecting the speed overrun.
Options
•
negative direction [-]
•
positive direction [+]
•
both directions [+/-]
Factory setting
•
both directions [+/-]
[MF2 Service] Laser
Warning
The output switches when the aging of the laser diode means that the
device is nearly due to be replaced.
Options
•
off
•
on
Factory setting
•
on
[MF2 Service] Level
Warning
The output switches when the signal attenuation value reaches the criti‐
cal value, for example due to contamination.
Options
•
off
•
on
Factory setting
•
on
[MF2 Service] Temp.
Warning
The output switches when the interior device temperature reaches the
critical value.
Options
•
off
•
on
Factory setting
•
on
[MF2 Service] Plausib.
Warning
The output switches if extrapolated measured values are to be output for
implausible measured values see table 21, page 44, ErrRej.
This may be caused by an internal error, hardware errors, obstruction of
the light beam, optical interference or electrical interference.
Options
•
off
•
on
Factory setting
•
on
10 PROFINET IO INTERFACE
72
O P E R A T I N G I N S T R U C T I O N S | DL100 Pro PROFINET HighFeature 8021823/19HC/2021-10-15 | SICK
Subject to change without notice