12
W.E.E.E Label - Found on individual parts of electrical system
(Waste Electrical and Electronic Equipment) Refer to section 15
Minimum patient weight
Temperature limit
Manufacturer
Humidity limit
Date of manufacture
Atmospheric pressure limit
Lot number
Minimum patient height
Minimum patient BMI
Class II electrical device
The user / occupant is protected by at least two layers of insulation between the current carrying parts
(e.g. control box and mains cable) and the metal accessible parts – If damage is noticed to any electrical
component, turn off at the mains and contact your provider or Sidhil Ltd. immediately.
Kneebreak to be used for lifting patient legs only
Type BF applied part
Applied Part: The parts of the bed that come into physical contact with the user / occupant in order for
the bed to carry out its intended function (refer to section 16.2 for a list of applied parts).
Type BF: Applied parts which are electrically isolated from earth and other parts of the medical equipment
- Complying with specific requirements for protection against electric shock to IEC 60601-1.
BMI 17
146 cm
40 kg