16.2 Electrical Data
Voltage in: 230V±10%, ~50/60Hz.
Current in: 1A
Mains cable fuse: 3-5A
Duty cycle:* 10%
2 mins of continuous use followed by 18 mins not in use.
*Electrically operated beds are intended to be operated intermittently rather than continuously. If the bed is operated
continuously for up to 2 minutes it must then be left for at least 18 minutes before reuse to allow the electrical system to cool
sufficiently. If the bed is continuously used for an extended period of time and it exceeds the duty cycle the control
box may become temporarily disabled or irreparably damaged.
Safety standards: IEC 60601-1
IEC 60601-11
IEC 60601-2-52
Electrical shock prevention:
Applied parts: Mattress platform
Profiling sections
Bed ends
Handset
Side rails
Liquid ingress protection: IPX4 - Splash resistant
Battery backup: Individual battery non replaceable
(SOLITE/PRO/BB & SOLITE/PRO/LOW/BB only) (Contact bed provider for battery part code)
Noise level: 66.9 dB(A) max
Service life: 10 years (battery life: 4 years)
Environmental conditions:
Operational Limits* Transportation / Storage Limits
Ambient Temperature +5ºC to +40°C -10°C to +50ºC
Humidity 20% - 90% at 30°C
non condensing
20% - 90% at 30°C
non condensing
Atmospheric Pressure 800 to 1060hPa 800 to 1060hPa
Altitude ≤2000m ≤2000m
* Always ensure the bed is brought up/down to room temperature before plugging in and operating -
Sidhil recommend the bed is left for a period of two hours or more to reach room temperature.