18.1 Mattress / Side Rail Compatibility Chart
GRANGE/SR SOLITE/PRO/SR SOLITE/PRO/SR/HE
Acclaim VE MAT/ACCL/VE/W
ü ü ü
Acclaim Profiler MAT/ACCL/PRO/W
ü ü ü
Softrest VE MAT/SOFT/VE
ü ü ü
Softrest Contour MAT/SOFT/CON
ü ü ü
Essential Foam MAT/BASIC
ü ü ü
Trio II DYN/DIG/TRIO/2
ü* ü* ü
Plus II DYN/DIG/PLUS/2
ü* ü* ü
SoloXtra DYN/DIG/SOLO/XTR
ü ü ü
Solo II
Softrest Thin
DYN/DIG/SOLO/2
MAT/SOFT/THIN
ü* ü* ü
Apollo DYN/DIG/1200
ü* ü* ü
Athena DYN/DIG/2000
ü* ü* ü
Acclaim Flow MAT/ACCL/FLOW
ü ü ü
Standard Grab Handle 1210/GRAB
ü** ü**
-
Padded Side 1316/PS/CREAM
ü
- -
Padded Side SOLITE/PRO/SR/PAD -
ü
-
Padded Side BED01/SR/HEXT/PAD - -
ü
Sidhil cannot be held responsible for any injury or incident which relates to the use of any product
combinations not approved by Sidhil Ltd.
It is the carer’s responsibility for selecting and fitting the products correctly and ensuring that the product
combination is compatible.
Warning
*
**
•
•
•
•
If the dynamic mattresses listed are used without side rail height extensions a patient
risk assessment must be performed to ensure the gap between the top of the
mattress and top of the side rail when raised is acceptable and will not introduce a
hazard to the patient.
Care must be taken when raising / lowering the GRANGE/SR and SOLITE/PRO/
SR when used in conjunction with the 1210/GRAB, as the gap between the two
components can introduce a finger entrapment hazard, positioning dependent.
If the dynamic mattresses listed are set to a low pressure the gap introduced by cell
compression at the mattress edge and the side rail is to be considered. A patient risk
assessment must be performed to ensure an asphyxiation risk is not introduced.
Ensure extension blocks are positioned centrally to the platform. If offset to one side
a gap will be introduced between the extension block and the side rail, introducing a
potential asphyxiation risk.
When the height extension is used in conjunction with a foam mattress ensure patient
ingress/egress is not compromised - If so, remove and refit the height extension as
required during patient transfers.
Care must be taken when raising the backrest when used in conjunction with the
1210/GRAB, as the gap between the two components can introduce a finger
entrapment hazard with the mattress retainers, positioning dependent.