Alternatively:
13.2 Steam Cleaning
The Solite Pro and Solite Pro Low can be dry steam cleaned. The individual
manufacturer’s instructions should be followed when using a steam cleaner and the
following precautions observed:
• Avoid directing steam directly at electrical components and reduce steam pressure
when cleaning near electrical items and connections.
• Avoid directing steam directly at wooden components and reduce steam pressure
when cleaning near wooden components.
• Use soft brushes and wiper cloths as recommended by the steam cleaner
manufacturer.
• Do not use excessive force or steam pressure on labels.
• Ensure the bed is dry and all debris from the cleaning process has been removed
prior to reuse.
• Ensure all electrical functions operate as normal once the bed has been cleaned
and dried.
Sidhil also recommend the use of Chlor-clean tablets. Follow the product
documentation for concentration guidelines and instructions for use.
Refer to the Sidhil infection control policy, copies are available from Sidhil Ltd. Contact
details can be found in section 2.
Sidhil recommend the use of Tristel ‘Fuse’ sachets and Tristel ‘Jet’ gel for cleaning of the
metal components. Follow the product documentation for concentration guidelines
and instructions for use.
Do not use high pressure hoses on the bed as they could cause damage
to the electrical components.
Caution
Caution
• The use of neat bleach or similar surface cleaners is not recommended
as damage may be caused to the cleaned surfaces.
• Sidhil do not recommend the cleaning of the wooden components
with Tristel products due to the risk of the solution leaching into the
wood.
FOR AUTHORISED PERSONNEL REFERENCE ONLY