36
Only authorised service personnel or Sidhil service engineers should carry out repairs
or service activities. Failure to do so may result in the manufacturer’s warranty
becoming void. The bed must be serviced once yearly, as a minimum.
14.3 Servicing
To maintain the bed’s essential performance the following checks must be performed:
Check that all electrical functions operate correctly on the handset.
Check that all electrical components and cables are in good condition - If not turn
off at the mains and remove bed from use until replacement parts are available.
Check the retaining clip is fastened to the control box, securing the electrical
cables in place.
Check that the red retaining ring is correctly inserted and clipped into the control
box securing the mains cable in place.
Warning
Failure to carry out the following checks at the stated frequency
could negatively influence the essential performance of the bed
and as a result put the patient at risk.
Always disconnect the bed from the mains power supply prior
to performing any maintenance procedures (when not checking
electrical functions).
Modification of the bed frame is not allowed without the
permission of Sidhil Ltd - A hazard could be introduced.
The bed should be vacated by the patient before any maintenance
or inspection takes place. If it is not possible due to the patient’s
mobility, a risk assessment should be carried out, and if deemed
safe to proceed, care should be taken for the service engineer to
avoid contact with the patient when working on electrical items.
Only Sidhil approved components, specified for the Solite Pro
and Solite Pro Low beds, should be used - if in doubt contact
Sidhil Ltd. or your local distributor.
Electrical system components are only to be replaced by
authorised service personnel or a Sidhil service engineer.
Never attempt to re-wire any components.
Linak battery packs may emit an increased amount of flammable
gas as they age - risk of explosion / fire. Sidhil advise that batteries
are replaced every 4 years or sooner (SOLITE/PRO/BB and
SOLITE/PRO/LOW/BB only).
FOR AUTHORISED PERSONNEL REFERENCE ONLY