Appendix A: Technical reference
A.4 Maximum Process Temperature Chart
SITRANS LR250 with mA/HART
244 Operating Instructions, 10/2019, A5E32220602-AG
A.4 Maximum Process Temperature Chart
WARNING
Exceeded maximum internal and process temperatures
Danger of device malfunction
• Internal temperature must not exceed +80 ° C (+176 °F).
• Process temperature must not exceed limits specified by the antenna type.
Note
•
The chart below is for guidance only.
•
The chart does not represent every possible process connection arrangement. For
example, it will NOT apply if you are mounting SITRANS LR250 directly on a metallic
vessel surface.
•
The chart does not take into consideration heating from direct sunlight exposure.
Maximum Process Temperatures versus allowable ambient
①
Internal enclosure temperature
②
Ambient temperature
③
Process temperature (at process connec-
tion)
Where the chart does not apply, please use your own judgement regarding the use of
SITRANS LR250.
If the internal temperature exceeds the maximum allowable limit, a sun shield or a longer
nozzle may be required.
See Current Internal Temperature (3.2.1.) to monitor the Internal Temperature.