• Attach the two limit switches so that the cam enables the switching mechanisms at the
switches (Fig. 12 / p. 20).
• When set correctly, you can hear the click emitted by the limit switches.
• Angulation 0°.
• Test the functioning of the limit switches by moving the C-arm into the +95° und -95° or-
bital positions.
• Check these settings at least twice.