Siemens AG SPR2-320.814.02.01.02 ARCADIS Orbic
01.05 CS SD 24
3D Reconstruction Upgrade Part 4 45
Page 45 of 46
Medical Solutions
7-
63D Reconstruction Upgrade Part 4
Checking C-arm movement for ARCADIS Orbic 3D 0
Proceed as follows after reconfiguring the basic unit and the monitor trolley.
• Connect the cables in the zipper hose to the basic unit.
• Connect the monitor trolley to the line voltage.
• Release the orbital brake.
• Manually move the C-arm into both end positions. The movement has to correspond to
"normal" C-arm movement. If not, verify that connector M3.X1 for the brake assembly is
inserted correctly. If yes, the LEDs in both keys light up.
The next step 0
Continue with ARCADIS Orbic Start-up Instructions SPR2-320.815.02... and Image Qual-
ity Assurance Test SPR2-320.820.02.. .
Perform a complete functional check.