26 3D Reconstruction Upgrade Part 1
ARCADIS Orbic SPR2-320.814.02.01.02 Siemens AG
01.05 CS SD 24
Page 26 of 46
Medical Solutions
Cable no. 4, item number n.a. 0
Use ribbon cable no. 4 to establish the connection D200.X203 - D1.X3.
Cable no. 5A, item number n.a. 0
Use ribbon cable no. 5A to establish connection D200.X212 - D30.X12.
Cable no. 5B, item number n.a. 0
Without the LITHOSTAR MODULARIS option:
Use ribbon cable no. 5B to establish connection D200.X211 - D30.X11.
Cable no. 2, item number n.a. 0
Disconnect plug D30.X4 on D30.
Insert extension cable no. 2 at plug D30X4A to D30.X4.
Insert plug D200.X114 on PC board D200.
Cable no. 1, item number n.a. 0
Connect cable no 1. to plug D30.X4 on D30 and to plug D200.X214 on D200.
Cable no. 6, item number n.a. 0
Connect cable no. 6 to plug D200.X220 on D200 and to D10.X12 on PC board D10.