4 Description
Hematek
®
3000 System – Instruction Manual – Version 1.3 114
4
Red waste tank light The red waste tank light is located at the right side of the slide stainer, next
to the rear conveyor spiral and below the yellow low stain light. It is off under
normal operating conditions.
A sensor activates the red waste tank light when the waste tank is full or
removed from the slide stainer. The waste tank should be emptied and
replaced.
Power module The power module is located on the rear of the slide stainer. The power
module is accessible from underneath and is displayed in Figure 4-12.
The power module converts the line current coming into the system to 24 V
direct current, which is the voltage within the system.
Main switch The main switch is located at the rear of the system and is accessible from
underneath. The main switch has two positions:
Position I: The power module is connected to power.
Position O: The power module is disconnected from power.
Service button The service button is located on the right side of the waste tank area and is
accessible after the removal of the waste tank. When the waste tank is
removed, a switch stops the pumps to prevent further draining of used
fluids. However, pressing the service button when the waste tank is
removed re-activates the pumps and allows fluid to be delivered to check
the stain-to-buffer ratio.
1 Power module
2 Socket
3 Main switch
4 Line fuse compartment
Figure 4-12 Power module
1 3 42