11
Siemens AG Healthcare Sector 11/12 Technical data
Multi-Nuclear Option
1
Multi-Nuclear Imaging and Spectroscopy for the following nuclei:
3He, 7Li, 13C, 17O, 19F, 23Na, 31P, 129Xe
Decoupling nuclei: 13C, 31P
Hyperpolarization nuclei: 3He, 13C, 129Xe
Frequency band 3He 7Li 13C 17O
Center frequency [MHz] 93.8524 47.8799 30.9806 16.7012
Usable transmitter band-
width [kHz]
±50 ±50 ±50 ±50
Usable receiver band-
width [kHz]
±190 ±100 ±65 ±50
19F 23Na 31P 129Xe
Center frequency [MHz] 115.9040 32.5885 49.8719 34.0816
Usable transmitter band-
width [kHz]
±50 ±50 ±50 ±50
Usable receiver band-
width [kHz]
±235 ±65 ±100 ±70
1. Optional