Siemens AG Healthcare Sector 11/12 Technical data
10
TimTX TrueForm
• Innovative techniques in the RF excitation hardware
• Uniform RF distribution in all body regions
TrueForm excitation uses amplitude and phase transmission settings optimized for
dedicated body regions. Feeding the 2 ports of the integrated body coil with an opti-
mized weighting yields a homogeneous B1 distribution
a-SPACE A version of the SPACE sequence. a-SPACE uses composite
adiabatic excitation pulses, which are insensitive to B1
spatial variations
B1 Filter An adaptive inline image filter that reduces any remnant
B1 effects without affecting image contrast
RF receiver technology
Number of coil elements Up to 204
Number of independent receiver chan-
nels
48, 64
1
, 128
1
Quadrature demodulation and filtering Digital
Receiver bandwidth 500 Hz–1 MHz (for each channel)
Receiver signal resolution 32 bit
ADC sampling rate 80 MHz
Preamplifier noise figure <0.5 dB
Dynamic range at coil connector
(referred to 1 Hz resolution bandwidth)
164 dB instantaneous at receiver
169 dB with automatic gain control at local
coil connector
1. Optional