Dose calculator
MAMMOMAT Balance SPB7-115.840.01 Page 42 of 62 Siemens AG
Troubleshooting Guide Rev.02 10.04 Medical Solutions
System configuration for calibration
• Compression paddle must be in
• Reference dosimeter must be placed under the compression paddle
• Dosimeter reference plane must be located at 550mm from focal spot
• Characteristics, response linearization, and unit of measure of reference Dosimeter must be
configured before to start calibration
• Reference dose outside kV range of reference Dosimeter must be set at ZERO
Radiation table
To view and edit the radiation table for the selected mammo
press the "Edit"-"Radiation table" menu.