Siemens MAMMOMAT Balance - Dose Calibration System Configuration; Radiation Table Management

Siemens MAMMOMAT Balance
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dose calculator
MAMMOMAT Balance SPB7-115.840.01 Page 42 of 62 Siemens AG
Troubleshooting Guide Rev.02 10.04 Medical Solutions
System configuration for calibration
Compression paddle must be in
Reference dosimeter must be placed under the compression paddle
Dosimeter reference plane must be located at 550mm from focal spot
Characteristics, response linearization, and unit of measure of reference Dosimeter must be
configured before to start calibration
Reference dose outside kV range of reference Dosimeter must be set at ZERO
Radiation table
To view and edit the radiation table for the selected mammo
press the "Edit"-"Radiation table" menu.