X-ray field alignment
MAMMOMAT Balance SPB7-115.840.01 Page 56 of 62 Siemens AG
Troubleshooting Guide Rev.02 10.04 Medical Solutions
Light / Beam alignment
Adjustment of the light field is necessary only if light beam collimator is disassembled.
To inspect light beam field use again a 24x30 table fully inserted as for normal use with a A3 format
paper and an 18x24 film on it, insert an 18x24 collimation plate.
Position 18x24 film as in the next picture to simulate x-ray field.
To correct front back edges of light field, adjust mirror inclination by means of screw .
SCREW