Gas spring replacement
MAMMOMAT Balance SPB7-115.840.01 Page 59 of 62 Siemens AG
Troubleshooting Guide Rev.02 10.04 Medical Solutions
3. unscrew gas spring from the base plate
4. replace gas spring assembly. Reassembling gas spring, take care of positioning sensing
microswitch shaft without damaging it.