Siemens AG SPR8-230.840.01.05.02 MOBILETT XP
11.05 CS SD 24
General information 9
Page 9 of 72
Medical Solutions
DANGER
To avoid electrical shock from components under voltage, also be
aware that:
The capacitors of the capacitor bank can be electrically charged
even when the system is switched off and the power cable is dis-
connected!
In the case of an error, individual capacitors of the capacitor bank
can still be electrically charged when disconnected from the
charging circuit!
Disconnect the battery blocks in the XP Hybrid/Digital prior to
maintenance and service work!
Carelessness can result in death or serious bodily injury.
¹ Make sure that no parts or tools fall into the unit;
¹ Do not touch potentially dangerous
components(Fig.2/p.10)
¹ If loose parts must be removed from the unit, use only
insulated tools;
¹ Protect the work area so that no other persons are able
touch the unit while the covers are open or removed!
• Switch the unit off before servicing or maintenance. Always disconnect the power plug
first.
• XP Hybrid: Make sure that the main switch and the battery mode are switched off.
• The capacitor bank discharges to < 40 V in approx. 15 minutes.
¹ The safety covers can be removed after this period has elapsed.
• Prior to performing any work, it must be verified that areas with dangerous voltage are
voltage-free (Fig.2/p.10).