10 General information
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 10 of 72
Medical Solutions
Fig. 2: Locations with dangerous voltage
Explanation of (Fig.2/p.10):
V Covers and safety covers are removed:
Back:
Left:
Right:
Additionally for XP Hybrid/XP Digital:
CAUTION: DC voltage (440V) at the capacitor bank (D927)
CAUTION: DC voltage (440V) directly at the capacitor bank!
CAUTION: DC voltage (440V) directly at the capacitor bank!
CAUTION: AC voltage (> 100V or > 60V ~ = half line voltage) at power
supplies U1 and U2 as long as the power cord is connected.
CAUTION: DC voltage (300 V) from the battery block to PCB D982!
Always disconnect the battery plug from BK1-BK4.