EasyManua.ls Logo

Siemens MULTIMOBIL 10 - Transformer Oil; Plastics; Electrolytic Capacitors; Safety Checks

Siemens MULTIMOBIL 10
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual
Service Instructions
Siemens Ltd. Med Division India Version 4.0
Page 10 of 48
Transformer Oil
Oil in X-ray tube assembly, approx. 7 kg.
Plastics
Epoxy resin on PC boards, PVC of cables approx. 10 kg.
Electrolytic Capacitors
These capacitors must be emptied. Please inform the Siemens Customer Service.
Capacitors approx. 15 kg
Remaining Electrolytic Capacitors approx. 0.5 kg.
1.21 Safety Checks
The user shall follow the following safety checks:
1.21.1 Daily checks
Check the power cord. Do not use the unit if the power cord is damaged.
Check the exposure indicator and audible signal during exposure.
Check that the Single tank assembly and the arm system remain in the desired
position.
1.21.2 Monthly checks
Check the lifting and lowering movements of the support arm.
Check function of locking lever and cap screw on the X-ray tube
Check the legibility of the labels according to those in section Location of Labels.
If any labels require to be replaced, please notify Siemens customer service.
1.21.3 Performance Check (Every Six months)
Check counterbalancing of the support arm at each position.
Check the locking mechanism for the support arm.
Check brake for proper functioning .
Take a sample exposure at 60 kV, 16 mAs.
Check for Radiation indication LED and the Audio indication in form of Buzzer.

Table of Contents

Other manuals for Siemens MULTIMOBIL 10

Related product manuals