3.1.2 Removing the Front cover.
3.1.3 Removing the X-ray source assembly.
Turn off the Multimobil 10,disconnect the mains chord and wind the cord around the
cable holder.
Lower the X-ray source assembly to transport position. Lock the Multimobil 10 by
applying the brake.
Make sure that the X-ray source assembly is properly secured with the transport
safety lock. It should not be possible to move the X-ray source assembly
upwards.This is very important for safety reasons.
Turn off the Multimobil 10,disconnect the
mains chord and wind the cord around the
cable holder.
Lower the X-ray source assembly to
transport position. Lock the Multimobil 10
by applying the brake.
Arrange some kind of soft bedding to place
the front cover on .
Unscrew the four ornamental screws with
2.5mm allen key.
Grasp the cassette box and slowly pull it
outside.
Place the front cover on a soft bedding.