Service Manual
Service Instructions
Siemens Ltd. Med Division India Version 4.0
Page 30 of 48
4.4 Initialisation codes
4.4.1 Code 90 (EPROM CHECKSUM FAILURE)
REPLACE J32 FIRMWARE
4.4.2 Code 96 (KV SOLL FAILURE)
ADJUST PRESET P7 ON D915
4.4.3 Code 97 ( mA FAILURE)
ADJUST PRESET P6 ON D915
START
CONFIRM 230v SUPPLY
SWITCH ON THE UNIT
ERROR CONTINUES
?
REPLACE D915 PCB
STOP
No
YES
IS
X15 CONNECTOR
SUPPLY OK ?
REPLACE D980
NO
Yes