4.6.10 Code 10 : Rotating Anode not OK
START
ALLOW 2 TO 10 SEC TIME IN BETWEEN
FIRST & SECOND STEP OF RADIOGRAPHY
Is anode in the
SingleTank rotating ?
No
Yes
Is ARR signal
obtained on D915?
No
Replace D915
Yes
Check feedback
voltages on test points
I1 and I2.
220V AC supply to
Rot. Anode ckt OK?
No
Yes
Check fuse F3 on D801
and replace if found
faulty.
If relay K2 on D801
doesn't operate,replace
K2
Replace D801
stop