EasyManua.ls Logo

Siemens MULTIMOBIL 10 - Code 15: JR < JRS

Siemens MULTIMOBIL 10
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual
Service Instructions
Siemens Ltd. Med Division India Version 4.0
Page 40 of 48
4.6.5 Code 15 : JR < JRS
START
CONFIRM mAS LINK ON
D800 PCB
SWITCH OFF THE UNIT
CHECK 'I' FOR 0.6V
+/ -0.1V AC RMS
Check for 22+/- 3 Ohms between
X8.5 and X8.6 and replace D800 if
found faulty.
KV build up OK?
No
Check Inverter ckt/
Inverter pulses/
Resonance cap. C5.
Yes
Voltage betw. X8.5
and X8.6 is OK?
No
Check and
replace D800 if
found fauly.
Check and replace
D915 if found faulty
Yes
STOP

Table of Contents

Other manuals for Siemens MULTIMOBIL 10

Related product manuals