EasyManua.ls Logo

Siemens MULTIMOBIL 10 - Code 06: JR <> 0

Siemens MULTIMOBIL 10
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual
Service Instructions
Siemens Ltd. Med Division India Version 4.0
Page 34 of 48
4.5.4 Code 06 : JR <> 0
START
MONITOR JR & SWITCH ON
THE UNIT
IS
JR = 0 & NO
SPIKES ?
REPLACE D915
ENSURE ALL CONNECTORS ARE
CONNECTED.
YES
STOP
R=20 OHM AT X8.5 &
X8.6
REPLACE D800.
IF THE ERROR CONTINUES,
REPLACE SINGLE TANK AND
CALIBRATE THE UNIT.
CHECK X8
CONNECTOR ON D915.
No
SWITCH OFF THE UNIT
No
Yes
Yes
REPLACE D915
STOP

Table of Contents

Other manuals for Siemens MULTIMOBIL 10

Related product manuals