POLYMOBIL III RXB8-115.034.02 Page 2 of 6 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
4 - 2 Functional checks
Checking the weight counterbalance 4
Unlock the support arm (1/Fig. 1) and loosen the knob (2/Fig. 1).
It should be possible to move the support arm easily without affixed accessories over the
entire range and it should come to rest in every position.
Readjustment of the weight counterbalance with affixed
accessories
4
Should the support arm not come to rest in every position with affixed accessories, then
readjust the spring tension for the weight counterbalance.
Remove the cover on the stand (3/Fig. 1).
Release completely the additional brake on the support arm (2/Fig. 1).
Move the support arm into the horizontal position.
Tighten the screw (4/Fig. 2) with a 17 mm fork wrench.
Adjust the spring tension so that there is equilibrium in the horizontal support arm
position, i.e. so that the forces for raising and lowering the tube assembly are the same.
Fig. 1 Fig. 2
1
2
3
4