5 - 1
Siemens AG RXB8-115.034.02 Page 1 of 2 POLYMOBIL III
Medical Solutions Rev. 03 04.05 CS PS 24
D
r
af
t
Final work steps 5
Protective conductor test 5
Observe the notes on the protective conductor test in these instructions.
Perform the protective conductor test while the system is closed according to
ARTD-002.731.17...
The protective conductor resistance may not exceed 0.2 ohms.
The values must be recorded as initial measurements in the report ’Protective conductor
resistance / report’ (chapter 7 of these instructions), stating the measuring points.
In addition document the measuring procedure and the measuring instrument used
(designation and serial number).
Device leakage current measurement 5
Observe the notes on the device leakage current measurement in these instructions.
Perform the device leakage current measurement while the system is closed according
to ARTD-002.731.17.
The limit of 2.5 mA must not be exceeded.
Electrical voltage!
Non-compliance can lead to severe injuries and even death.
In the direct measurement of the device leakage current no hous-
ing parts of the system may be touched during the measurement.
Third-person access to the system must be prevented.
Fig. 1
WARNING