POLYMOBIL III RXB8-115.034.02 Page 2 of 8 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
1 - 2 Requirements
Required tools and measurement devices 1
Standard service tool kit*
Ground wire tester
Device leakage current measuring device
Digital multimeter
Safety information and protective measures 1
All tools, measuring equipment and aids, with the exception of
those marked with "*", are listed and specified in the STC
(Service Tools Catalogue).
Danger of injuries, death or material damage.
Note
- the product-specific safety notes in these instructions,
- the general safety information in the document
TD00-000.860.01...
and
- the safety information in accordance with ARTD Part 2.
Non-compliance can lead to death, to injuries or to
material damage.
X-ray radiation!
Non-compliance can lead to illnesses, irreversible damage to
body cells and the genotype, severe injuries and even death.
In work on the system in which radiation must be released, the
radiation protection directives and the rules for radiation protec-
tion according to ARTD 02.771.02 must be complied with.
Please note:
- Use available radiation protection devices.
- Wear radiation protection clothing (lead apron).
- Keep your distance from the radiation source as large as possible.
- Release radiation only if necessary.
- Set radiation activity as low as possible.
(Low kV and mAs values)
- Release radiation for as short a time as possible.
Checks in which radiation must be released are identified by the
radiation warning symbol.
NOTE
WARNING
WARNING