POLYMOBIL III RXB8-115.034.02 Page 4 of 6 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
7 - 4 Electrical safety / reports
Measuring circuit
Direct measurement.
Differential measurement
Fig. 2 Measuring circuit for the direct measurement of the device leakage
current according to DIN VDE 0751-1:2001-10, Fig. C5
for protective class
I.
1 = System
2 = Application part (if present)
3 = Measuring arrangement (integrated in the measuring
instrument)
Fig. 3 Measuring circuit for the measurement of the device leakage
current according to the differential current method
complying with DIN VDE 0751-1:2001-10, Fig. C6
for protective class
I.
1 = System
2 = Application part (if present)
3 = Measuring arrangement (integrated in the measuring
instrument)