28 SFIDK.PS.029.Q3.02
Signal conv. inst.
5. Signal converter installation
SITRANS F US SONOFLO
®®
®®
®
5.12.5 Electrical connec-
tion for "flame-
proof housing"
compact version
5.12.6 Electrical connec-
tion diagram
compact version
Supply voltage 230 V a.c.
115 to 230 V a.c. is connected to terminals 1 and 2. The ground wire must be connected to the
ground terminal on the terminal plate.
If the ground wire is not connected, personnel can be exposed to 115 V / 230 V.
- this page has been updated 2003.11