SFIDK.PS.029.Q3.02 9
2. Technical data
Tech. data
SITRANS F US SONOFLO
®®
®®
®
Introduction 1-track or 2-tracks sensor supplied as a kit for mounting on
existing pipelines
Description DN 200 to DN 4000 (8" to 160")
Line pressure Max. 40 bar
Liquid temperature Standard version: −20 °C to +200 °C
Enclosure Standard version: IP 68
Ex version: EEx d IIC T3-T6, IP 68
Materials Standard version: PA 6.6 100 °C
Terminal box AISI 316 200 °C
Transducer element Standard version: AISI 316 200 °C
Material of existing pipeline
Steel Transducer holder: St. 37.2 or simular
Mounting plates: St. 37.2 or simular
Concrete Transducer holder: ASTM 316 or simular
Mounting plates: (Not included)
Stainless steel Transducer holder: ASTM 316 or simular
Mounting plates: ASTM 316 or simular
Pipe wall thickness
Steel pipe (AISI 316 and St. 37.2 or Transducer and holder available in length L = 160
corresponding material) Allowing a pipewall thickness up to 20 mm
Concrete pipe Transducer available in lengths L = 230
Allowing a pipewall thickness up to 200 mm
2.5 SONOKIT mounting
kit
Coaxial cable (75
ΩΩ
ΩΩ
Ω) 1 x ∅ 0.8 mm copper conductor with shield
Outside diameter ∅ 5.8 mm
Length Max. 250 m between sensor and signal converter
Material (outside jacket) Black PE
Ambient temperature −10 °C to +100 °C
Coaxial cable (75
ΩΩ
ΩΩ
Ω) 1 x ∅ 0.65 mm copper conductor with shield
Outside diameter ∅ 5.3 mm
Length Max. 100 m between sensor and signal converter
Material (outside jacket) Brown FEP
Ambient temperature −200 °C to +200 °C
Multi coaxial cable (75
ΩΩ
ΩΩ
Ω) 4 x coax
Outside diameter 9 mm
Length Max. 20 m
Material (outside jacket) Black PVC
Ambient temperature 0 °C to +70 °C
Multi coax cable (75
ΩΩ
ΩΩ
Ω) 4 x ∅ 0.30 mm copper conductor with shield and joint jacket
Outside diameter ∅ 7.9 mm
Length Max. 10 m
Material (outside jacket) Black PTFE
Ambient temperature −50 °C to +200 °C
2.6 Coaxial cable for
SONO 3300, 3100/
3000