Operation
7.3 Menu system
SITRANS SL
118 Operating Instructions, 12/2010, A5E01132948-04
Digital outputs
The digital outputs are realized as relays providing the following functions:
Assign alarm(s) from the alarm list to a digital output. Possible settings are:
M(aintenance) fault
M demanded
M required
P(rocess) V(alue) fault
PV warning
PV tolerance
No data exchange
Local override
Configuration warning
Data exchange
Device unlocked
To assign or cancel an alarm use the
(right arrow) button.
Select state of relay in normal operation
"Normally energized" - relay contact for the alarm signal is closed.
"Normally de-energized" - relay contact for the alarm signal is open.
"Normal" refers to the normal state of operation of the analyzer. The system is powered and
running without any problems, no error signal is generated and displayed.
"Normally Energized" refers to the state of the relay in normal operation. The relay contact
for the alarm signal is closed in normal state of operation when checked.