Installing
4.3 Electric connections non Ex
SITRANS SL
48 Operating Instructions, 12/2010, A5E01132948-04
4.3.2 Screw terminals
Receiver housing
CAUTION
Damage to the equipment
Damage to the equipment may result from short-circuit. No conductors of a cable may be
left unconnected inside the SITRANS SL receiver or transmitter. Any unused conductor
must be connected to the analyzer housing.
NOTICE
No Ethernet connection possible
If Ethernet is connected to a permanently installed cable, the Ethernet jack inside the
receiver may not be used. Before using the Ethernet jack inside the receiver, ensure that no
cable conductors are connected to the Ethernet screw terminal in the receiver.