2.4 Design
SITRANS SL
16 Operating Instructions, 12/2010, A5E01132948-04
2.4 Design
The SITRANS SL gas analyzer consists of a pair of cross-duct sensors - a transmitter unit
and a receiver unit - both with the same dimensions. The complete analyzer is integrated in
these two enclosures (pos. 1 and 7, fig. 2.2). The transmitter unit contains the laser source
which light is transmitted to the receiver through the measurement path. The receiver unit
contains a photodetector including electronics as well as a reference cell. The receiver unit is
connected to the transmitter unit by means of a sensor connection cable. The connection
cable of the receiver is used to connect the power supply and the communication interfaces.
The receiver enclosure contains a local user interface (LUI) with an LC display which can be
read through a window in the lid. The LUI is normally operated via a remote control panel.
2.4.1 Transmitter and receiver units
Special features of the transmitter and receiver units:
Two in situ cross duct sensors, designed as transmitter and receiver units, connected via
sensor connection cable
Process interface which
consists of identical assemblies for transmitter and receiver
is the interconnection between the transmitter/receiver units and the process
Powder-coated aluminium; stainless steel
Degree of protection IP65
Adjustable flanges
Suitable for a flange connection to DN50/PN10-40, ANSI 4"/150 lbs
Purging facilities on the process and the sensor sides, configurable application with
purging gas connections