4.3 Electric connections non Ex
SITRANS SL
46 Operating Instructions, 12/2010, A5E01132948-04
4.3 Electric connections non Ex
EMC and NAMUR
The following must be observed regarding EMC and NAMUR:
CAUTION
This is a Class A product. In a domestic environment this product may cause radio
interferences in which case the user may be required to take adequate remedial measures
Operating SITRANS SL
Since the device is delivered without cabling the user must provide this. An appropriate cable
layout is described in section 3.1.
The cables have to be connected according to the table in section 4.5.3.
The following must be observed when operating SITRANS SL:
WARNING
Never switch on or operate an analyzer with the lid open.
The protective earth terminal of the SITRANS SL needs to be connected to the local
potential equalization system.
See also
Connecting cables (Page 52)
Cable selection recommendations (Page 31)