Alarm, error, and system messages
10.3 PROFIBUS diagnosis and alarms
SITRANS SL
Operating Instructions, 12/2010, A5E01132948-04
179
Bit position Error text message in GSD file Error Code
Unit_Diag_Bit(53) "Prs. comp. file range" E30
Unit_Diag_Bit(54) "CurveFit temp. file range" E31
Unit_Diag_Bit(55) "CurveFit prs. file range" E32
Unit_Diag_Bit(56) "CurveFit gas file range" E33
Unit_Diag_Bit(57) "NH
3
outside range" E34
Unit_Diag_Bit(58) "O
2
outside range" E35
Unit_Diag_Bit(59) "H
2
O outside range" E36
Unit_Diag_Bit(60) "HF outside range" E37
Unit_Diag_Bit(61) "H
2
S outside range" E38
Unit_Diag_Bit(62) "HCl outside range" E39
Unit_Diag_Bit(63) "HCN outside range" E40
Unit_Diag_Bit(64) "CO outside range" E41
Unit_Diag_Bit(65) "CO
2
outside range" E42
Unit_Diag_Bit(66) "Temperature outside range" E43
Unit_Diag_Bit(67) "CH
4
outside range" E44
Unit_Diag_Bit(68) "FPGA timeout receiver" E45
Unit_Diag_Bit(69) "FPGA timeout receiver junction" E46
Unit_Diag_Bit(70) "FPGA timeout transmitter" E47
Unit_Diag_Bit(71) "Laser unpowered " E48
Unit_Diag_Bit(72) "Transmitter board temperature" E49
Unit_Diag_Bit(73) "Receiver board temperature" E50
Unit_Diag_Bit(74) "Laser board temperature" E51
Unit_Diag_Bit(75) "Detector board temperature" E52
Unit_Diag_Bit(76) "Receiver junction board temp." E53
Unit_Diag_Bit(77) "Laser vibrator temperature" E54
Unit_Diag_Bit(78) "Analog output test mode" E55
Unit_Diag_Bit(79) "Digital output test mode" E56
Unit_Diag_Bit(80) "EEPROM configuration missing" E57
Unit_Diag_Bit(81) "Save settings failed" E58
Unit_Diag_Bit(82) "Sensor temp. comp. file range" E59
Unit_Diag_Bit(83) "Watchdog reset triggered" E60
Unit_Diag_Bit(84) "LUI unlocked" E61
Unit_Diag_Bit(85) "Hardware inconsistency" E62
Unit_Diag_Bit(86) "A0 outside measurement range" E63
Unit_Diag_Bit(87) "A1 outside measurement range" E64
Unit_Diag_Bit(88) "Laser drift" E65
Unit_Diag_Bit(89) "Algorithm error" E66
Unit_Diag_Bit(90) "Measurement quality" E67
Unit_Diag_Bit(91) "Line locking failure" E68