SSA3000X User Manual 72
3.3.2 Send SCPI Commands via NI MAX
Users can control the spectrum analyzer remotely by sending SCPI commands via NI-MAX
software.
3.3.2.1 Using USB
Run NI MAX software.
1, Click “Device and interface” at the upper left corner of the software;
2, Find the “USBTMC” device symbol
3, Click “Open VISA Test Panel” option button, then the following interface will appear.
4, Click the “Input/Output” option button and click the “Query” option button in order to view
the operation information.
NOTE: The *IDN? command (known as the Identification Query) returns the instrument
manufacturer, instrument model, serial number, and other identification information.
3.3.2.2 Using LAN
select, Add Network Device, and select VISA TCP/IP Resource as shown:.
Run NI MAX software.
1, Click “Device and interface” at the upper left corner of the software;
2, Find the “Network Devices” symbol , click “Add Network Devices”;