Safety Protection Device and Troubleshooting
7-2
7.4 Failure Analysis and Troubleshooting
Failure Diagnosis
The device has a failure self-test function. If any key failure occurs, the device will give
alarm, which facilitates failure inspection and troubleshooting.
Generally, the first step to confirm a failure is to check if the voltage of the power supply
is normal, and check all circuit boards for any failure.
Key points to determine failures
If any device error occurs, the component failed can be determined from the displayed
error. Check the test points of the relevant components and signal indicators to quickly
determine the cause of failure, so as to restore the device to normal in time.
Error message
When a system error occurs, the portable X-ray unit sounds "di- di" and an error message
is displayed on the control panel. The corresponding content and troubleshooting of the
error prompts are described in Table 7-1.
Table 7-1 Error Prompt List
Error Code
Failure Description and Cause
Analysis
Troubleshooting
Oiltank
Overheat
Oiltank Overheat
Click the error display area to clear
the error and retest. If the error still
exists, restart the system after device
cooling.
ERROR kV
KVFB voltage value
in standby state
Click the error display area to clear
the error and retest. If the error is still
reported, shut down and contact the
manufacturer for service.
ERROR mA
MAFB voltage value
in standby state
Click the error display area to clear
the error and retest. If the error is still
reported, shut down and contact the
manufacturer for repair.
ERROR kV IN
EXP
No tube voltage output
in exposure.
If the failure reproduces, contact the
manufacturer for repair.