Guidance and Manufacturer’s Declaration
B-1
Appendix B
Guidance and Manufacturer’s Declaration – Electromagnetic
Emissions for All Equipment and Systems
To provide reasonable protection, this product complies with emissions limits for a
Group
1, Class A ISM equipment Medical Devices Directive as stated in EN 60601-1-2.
【
Warning 1】: The use of accessories, sensors, and/ or cables not supplied by SIUI may
lead to increase of electromagnetic radiation or decrease of electromagnetic immunity,
consequently resulting in operation failure.
【Warning 2】: The distance between any part of SR-8230S (including cables) and
portable RF communication equipment shall not be less than 30cm, otherwise it may
result in declined device performance.
【Warning 3】: Precaution to be taken to prevent adverse events to the PATIENT and
OPERATOR due to ELECTROMAGNETIC DISTURBANCES is to ensure that the main unit is
grounded reliably.
【Note】: The EMISSIONS characteristics of SR-8230S make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment might not
offer adequate protection to radio-frequency communication services. The user might
need to take mitigation measures, such as relocating or re-orienting the equipment.