Safety Protection Device and Troubleshooting
7-3
LOW BATTERY Low battery
Click the error display area to clear
the error and retest. If the error is still
reported, recharge the battery.
HIGH BATTERY High battery
Click the error display area to clear
the error and retest. If the error is still
reported, shut down and contact the
manufacturer for repair.
HET Exposure interval time limit
It would be automatically clear when
the exposure interval was over.
Remark: If the failure cannot be fixed by following the troubleshooting instruction above,
please contact the factory for service.
【
】
If the display on the control panel is in disorder, the display
parameters go beyond normal range, or any other unpredictable
abnormal situation occurs, DO NOT continue the operation. Cut off
the power supply of the main unit immediately. Restart the device
after 1
minute, and the device may return to normal usually. If the
display disorder still exists, contact the factory for service.
Note: The error prompt list may be longer or shorter due to software
update. For specific error prompts and the failure description, see the
software version description. The error prompt list and the failure
description are updated without notice.
7.5 Replaceable components list
Table 7-2 Replaceable part List
1 F2 Blade fuse 257040 4 LittleFuse
Instruction for component replacement:
Replacement of fuse: Remove the upper enclosure of the portable X-ray machine, and
insert the fuse that complies with the model and parameter requirements in the part list
above into the corresponding position.