PRODUCT SPECIFICATIONS
SKF Multilog On-line System IMx-16Plus
SKF Multilog On-line System IMx-16Plus
User Manual
Revision Letter A
3.2.7 Digital measurement
Frequency range: 0,016 Hz to 20 kHz (1 cpm to 1,2 Mcpm)
o Note that when used for order tracking, the maximum pulse frequency
is 2,5 kHz
Frequency accuracy: 0,05% of measurement value
(typically 0,01% up to 2,5 kHz)
Pulse counting
Configurable pulses per rev.
o The product of pulses per rev and rotational speed is subject to the
maximum frequency range, limitation
3.2.8 Signal processing
Event capture mode
o Data storage on time, associated measurement value or alarm state
o Run cycle mode (repeated or cyclic captures)
Time waveform
Vector analysis with circular alarms
SKF enveloping filters I, II, III, IV
FFT: 100 to 6 400 lines
o Integration/Differentiation in the frequency domain
o Window function: Hanning
Customer formulated mathematical equations
Dynamic alarm levels, active range determined on multiple parameters
Data storage on time, event or alarm condition
Detection of sensor and cable fault
Watchdog and self-testing
3.2.9 Interfaces
The mobile data and Ethernet interfaces are alternatives and cannot be used
concurrently
Mobile data
o Global GPRS and LTE (2G/3G/4G networks)
o Auto switching between network types
o Dedicated antenna connections provided
LTE 1/2 SMA female
o DNS lookup and NTP time synchronisation are supported
o Firmware configurable support for micro-SIM or eSIM.
Micro-SIM card holder (top end cap)
Ethernet:
o RJ45,
100 Mbit/s TCP/IP (one port)
Auto cable detection
Support for PoE, refer 3.2.2 and 2.5.3
o Wi-Fi
802.11n
2,4 GHz band, only
Open or secured networks, WPA2-PSK
Dedicated antenna connection provided: Wi-Fi SMA female

Table of Contents