Page 6
1-3. Preventive Maintenance Procedures
The following procedures should be done semi-
annually or sooner as needed.
a. Lighthead top cover should be removed and
any accumulation of dust or lint removed.
b. Check the focus drive mechanism for proper
operation and lubrication. Lubricate the drive
gears with lithium grease and all pivot points with
a lightweight oil as necessary.
c. All attaching hardware (screws, nuts, etc.)
should be physically checked for tightness. Any
missing hardware MUST be replaced.
WARNING
Apply LOC-TITE to any replacement
or loose vertical support tube attach-
ing screws.
d. Rotate the radial arm assemblies around the
ceiling mount to check the slip ring/brush block as-
semblies. With electrical power "ON", if the lights
become intermittent or go out, check and repair or
replace the brush block or slip ring as necessary.
e. Check the remaining slip ring/brush block
assemblies by raising, lowering, and rotating the
lighthead through its full range of motion. With
electrical power "ON", if the lights become inter-
mittent or go out, check and repair or replace the
appropriate brush block or slip ring as necessary.
f. Check all fixture rotation axes for proper ad-
justment. Adjust as necessary using applicable
adjustment procedures from the adjustment section
of this manual (Section 2).
SECTION I MAINTENANCE
1-1. General
To ensure proper operation and extend the life of
your SKYTRON surgical lighting fixture, the fol-
lowing preventive maintenance procedures are
recommended.
NOTE
All repairs should be made using autho-
rized SKYTRON replacement parts.
1-2. Daily Maintenance
Daily or between cases, the lighthead exterior
should be wiped down with a mild cleaning agent
which will not affect the painted or acrylic parts.
Do not apply or spray cleaning agents directly on
the lighthead.
CAUTION
Use of incompatible cleaning agents will
cause damage to the fixture. Avoid the
use of cleaning solutions which contain
high concentrations of alcohol, ethelene
glycol, phenol, iodophors, or glutaral-
dehyde based disinfectancts. Staining,
pitting, discoloration and diffuser crack-
ing may occur if these are used.
•Onlydissinfectantsandcleanersmaybeusedthat
are certified from the manufacturer for compatibility
with the following materials:
-- Polycarbonate (PC)
-- Polyvinylchloride (PVC)
-- Silicones
•Alwaysconsultwiththemanufactureroftheclean-
ing agent for proper application and use. Always
spot test on an inconspicuous area before use.
•Avoidpersonalinjury.Donotattempttocleanlight-
head unless power is turned off at wall control.
•Avoidusingexcessiveamountsofspraycleaners.
Leakage of fluids into the interior of lighthead may
cause corrosion of electrical components.