Page 18
Handle Cleaning & Sterilization
The sterilizable handles are constructed of heat
resistant, impact resistant plastic. They should
be cleaned with mild alkaline cleaning products
WITHOUTactivechlorine.Thoroughlyrinseoffall
cleaners with water. Ensure that the open side of
handles are face down.
Recommended sterilization parameters for steriliz-
able handle:
Steam Sterilization:
134°C(273°F):2.3bars(33.3PSI)
4 minutes
Always consult current AORN journal recommen-
dations for proper sterilization procedures.
Adjustments
As part of a regular preventive maintenance pro-
gram, it is suggested that a check of the various
positioning axes be made to verify correctness of
tension adjustment. If any lighthead drift is noticed,
all that is usually necessary is a minor adjustment.
Readjustment should be made as per the appro-
priate instructions contained in the Maintenance
and Adjustment Manual. Also, during a scheduled
cleaning of the lighthead interior, lubrication of the
various moving parts is desirable.
MAINTENANCE
General
To insure proper operation and to extend the life
of your SKYTRON surgical lighting system, the
following preventive maintenance procedures are
recommended.
NOTE
All repairs should be made using only
authorized SKYTRON replacement
parts.
Daily Maintenance
Daily or between cases, the lighthead exterior
should be wiped down with a mild cleaning agent
which will not affect the painted or acrylic parts.
Do not apply or spray cleaning agents directly on
the lighthead.
CAUTION
Use of incompatible cleaning agents
will cause damage to the fixture. Avoid
the use of cleaning solutions which
contain high concentrations of alcohol,
ethelene glycol, phenol, iodophors, or
glutaraldehyde based disinfectancts.
Staining, pitting, discoloration and dif-
fuser cracking may occur if these are
used.
•Onlydissinfectantsandcleanersmaybeusedthat
are certified from the manufacturer for compatibility
with the following materials:
-- Polycarbonate (PC)
-- Polyvinylchloride (PVC)
-- Silicones
•Alwaysconsultwiththemanufactureroftheclean-
ing agent for proper application and use. Always
spot test on an inconspicuous area before use.
•Avoidpersonalinjury.Donotattempttocleanlight-
head unless power is turned off at wall control.
•Avoidusingexcessiveamountsofspraycleaners.
Leakage of fluids into the interior of lighthead may
cause corrosion of electrical components.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Skytron AURORA LED Series and is the answer not in the manual?

Skytron AURORA LED Series Specifications

General IconGeneral
BrandSkytron
ModelAURORA LED Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals