WARMING CABINET OWNER’S
• REV2
4-2-5. Notications
Current settings, operational conditions and the
current logged in user can be quickly viewed.
Touch: (1)yellow time/date to open a drop-down
list. Swipe (2)down-up to view the whole list. Touch
(1)time/date to close the list (Figure 20).
MM/DD/YYYY
HH:MM AMPM
Notifications
Single 1
PV = 25.0 SP = 30.0
Program
Inactive
Alarm
Inactive
Data
1
2
Figure 20. Notications List
4-2-6. Temperature Setpoints
All warming chambers setpoint temperature range
are preset to a minimum 90° F (32 °C) and maximum
160° F (71 °C).
Temperature settings may be changed at any time.
An alarm will activate if the setpoint is changed
more than 10° below the actual temperature.
All loads should be allowed time to stabilize at the
Setpoint temperature:
Solutions - Approximately 8 to 12 hours
Blankets - Approximately 6 to 8 hours
Temperature settings for uids vary depending
on use or manufacturer. Always refer to
AORN guidelines and the uid manufacturer’s
recommendations for proper temperature settings.
CAUTION
Burn Hazard: DO NOT raise the setpoint
to increase heating rate. On dual chamber
models, do not exceed 30° F (16.7 °C)
higher than an upper chamber. Example:
Upper chamber 110° F (43 °C) lower
chamber must not exceed 140° F (60 °C).
This could overheat contents leading to
possible patient burns.
WARNING
Explosion Hazard: DO NOT exceed 150°
F (66 °C) for non-vented closures; screw
caps, crimp seals, plastic pouches, etc.
DO NOT exceed pre-sterile solution
manufacturer’s temperature requirements.
Home screen shows current temperature - green,
setpoint - red, output percent - white (Figure 21 &
Figure 22).
MM/DD/YYYY
HH:MM AMPM
Single
F
100.0
%
150 SP
Figure 21. Single Chamber Home Screen
MM/DD/YYYY
HH:MM AMPM
Upper
F
100.0
%
150 SP
Lower
100.0%
150 SP
F
Figure 22. Dual Chamber Home Screen
Change setpoint: touch red set point(s) to open
keypad (Figure 23). Key new set point, touch Done.
New set point appears on screen.
SINGLE (
UPPER or LOWER
) SP
<-- Clear
Cancel Done
1 2 3
4 5 6
7 8 9
+ - 0
Min: 90
Max: 160
Figure 23. Setpoint Keypad