Page 5
WARMING CABINET OWNER’S
• REV2
2-1. Intended Use
CAUTION
Explosion or Fire Hazard: The Warming
Cabinet is intended for warming
non-ammable solutions and blankets
made of Muslin, 100% cotton or wool,
ONLY. Operating the Warming Cabinet
with ammable items could result in
explosion or re.
The SKYTRON Warming Cabinets are designed to
store and warm blankets, hospital linens, irrigation
uids, and/or injection uids in accordance with
recommended warming temperatures and storage
time guidelines provided by the manufacturers of
such products.
SECTION 2. EQUIPMENT SPECIFICATIONS
2-2. Recommended Settings
SKYTRON does not recommend chamber set points
for any items that are to be warmed. For appropriate
heating temperatures, please contact the item
manufacturers. For more information, please
contact your SKYTRON service representative.
For blankets, follow blanket manufacturer’s
instructions for the set point.
For intravenous and irrigation fluids, follow
temperature guidelines printed on the container or
contact your supplier for temperature and expiration
periods.
2-3. Installation
The SKYTRON Set-up Instructions specify the
unpacking, installation, and testing of the Warming
Cabinet. Prior to starting the installation, review
the Set-up Instructions and local electric codes
including the Occupational Health and Safety
Act for any requirements pertaining to the proper
installation of this equipment.
Contact your SKYTRON representative for seismic
calculations and tie-down hardware if applicable.
2-4. Environmental Conditions
DURING TRANSPORT AND STORAGE (IN ORIGINAL PACKAGING MATERIALS)
• AMBIENT TEMPERATURE: -40° - 159° F (-40° - 70° C)
• RELATIVE AIR MOISTURE: 10% - 100%, INCLUDING CONDENSATION
• AIR PRESSURE: 14 inHg - 31.3 inHg (500 hPa - 1060 hPa)
DURING USE - FOR DRY LOCATIONS
• AMBIENT TEMPERATURE: 60° - 85° F (15° - 30° C)
• RELATIVE AIR MOISTURE: 30% - 60% NON CONDENSING
• AIR PRESSURE: 20.7 inHg - 31.3 inHg (700 hPa - 1060 hPa)
The Warming Cabinet is intended for use in a stable environment, with an ideal ambient temperature of 72°
F (22°C) or less. The Warming Cabinet should never be located next to any appliance that may produce
heat, such as an autoclave. If the Warming Cabinet must be located next to a heat producing appliance,
action must be taken to prevent heat transfer.
2-5. Certication