Messages and Alarms
When you see and hear: Take this action:
Two-Tone alarm
No Disposable,
Clamp Tubing
The disposable (CADD
®
administration set or
CADD
™
medication cassette
reservoir) has been removed
or may have become
depleted, or the disposable is
not aligned with the pump or
is damaged, or a malfunction
of the pump sensor(s) is
occurring. Clamp the tubing
immediately. A disposable
must be properly attached
in order for the pump to
run. Press
⁄
or
„
to
silence the alarm. If the
alarm persists, contact your
clinician.
Two-Tone alarm
No Disposable,
Pump won't run
You have tried to start the
pump without a disposable
(CADD
®
administration set or
CADD
™
medication cassette
reservoir) attached, or the
disposable is attached but is
not aligned with the pump or
is damaged, or a malfunction
of the pump sensor(s) is
occurring. A disposable must
be properly attached in order
for the pump to run. Press
⁄
or
„
to silence the
alarm. If the alarm persists,
contact your clinician.