Messages and Alarms
When you see and hear: Take this action:
Two-Tone alarm
Power lost while
pump was on
The pump was running when
power was removed. Stop
the pump before changing
the batteries or removing the
power source. Press
or
to silence the alarm.
Two-Tone alarm
Reservoir Volume
Empty
The Reservoir Volume has
reached 0.0 mL. Press
or
to silence the alarm.
Then install a new fluid
container if appropriate and
reset the reservoir volume.
Three single beeps
RUN
ResVol Low
The Reservoir Volume is low.
Change the fluid container
soon. See page 16 for
further details.
Two-Tone alarm
Service Due
See manual
The pump is scheduled for
service. Press
or
to
silence the alarm. The pump
is still functional, but contact
your clinician for further
instructions.