Messages and Alarms
When you see and hear: Take this action:
Two-Tone alarm
Upstream Occlusion
Fluid is not flowing from the
fluid container to the pump.
Check for a kink in the tubing
or a closed clamp between
fluid container and pump.
Press
or
to stop
the pump and silence the
alarm for 2 minutes. After
you remove the cause of
upstream occlusion, start
the pump if necessary. If the
alarm continues, contact your
clinician.