USE OF ADAPTORS, EXTENSIONS AND UNIVERSALS
Anytime an adaptor, extension or universal is used with a torque wrench in such a way that
fastener distance is different than torque wrench square drive distance at calibration, an
adjustment to head length is required to get a proper fastener torque reading.
When using wobble extension or a universal, do not exceed more than 15 degrees of offset from
perpendicular drive. Do not use a long extension with flex-drive at full flex.
CALIBRATION
Contact your Snap-on sales representative for calibration services or refer to ControlTech
TM
Calibration Manual.
CERTIFICATION
This torque-angle wrench was calibrated at factory using angular displacement and torque
measurement instruments that are traceable to National Institute of Standards and Technology
(N.I.S.T.). Torque parameters comply with ISO 6789:2003 and ASME B107:300-2010 (B107.29).
Note: no U.S. or International Standards exist for angle wrenches.
MAINTENANCE / SERVICE
Clean wrench by wiping with a damp cloth. DO NOT use solvents, thinners or carburetor
cleaners. DO NOT immerse in anything.
Service and repairs are to be done by Snap-on Service Centers only. Contact your Snap-on Tools
representative.
Ratchet head repair kits can be ordered from a Snap-on Representative.
NOTES: - If display shows persistent "TORQUE ZERO ERROR" at power on, wrench is damaged
and must be returned for repair
- If display shows ANGLE ERRORin angle mode, fastener rotation speed has
exceeded capacity of wrench.
- Wrench must be held still during angle zeroing. Motion is indicated by alternating
dashes - - on display
- Remove battery when stored for extended periods (Note: clock will revert to default
settings).
IMPORTANT! Calibration events are recorded in wrench memory
which provides evidence to void factory certification.