MPI-520 – USER MANUAL
Measurement of resistance with low current
Voltage at open terminals: 4…9 V
Output current < 8 mA
Audio signal for resistance being measured < 30 ±50%
Compensation of test leads resistance
Measurement of insulation resistance
Test range according to IEC 61557-2 for U
N
= 50 V: 50 k…250 M
Display range for
U
N
= 50 V
(3 % w.m. + 8 digits),
[ (5 % w.m. + 8 digits)] *
* - for WS-03 and WS-04 leads
Test range according to IEC 61557-2 for U
N
= 100 V: 100 k…500 M
Display range for
U
N
= 100 V
(3 % w.m. + 8 digits)
[ (5 % w.m. + 8 digits)] *
* - for WS-03 and WS-04 leads
Test range according to IEC 61557-2 for U
N
= 250 V: 250 k…999 M
Display range for
U
N
= 250 V
(3 % w.m. + 8 digits)
[ (5 % w.m. + 8 digits)] *
* - for WS-03 and WS-04 leads
Test range according to IEC 61557-2 for U
N
= 500 V: 500 k…2.00 G
Display range for
U
N
= 500 V
(3 % w.m. + 8 digits)
[ (5 % w.m. + 8 digits)] *
(4 % w.m. + 6 digits)
[ (6 % w.m. + 6 digits)] *
* - for WS-03 and WS-04 leads