MPI-520 – USER MANUAL
t) memory of measurement results ............................................................... 990 cells, 57,500 results
u) data transmission .................................................................................................................... USB
v) quality standarddevelopment, design and manufacturing are ISO 9001, ISO 14001, PN-N-18001
compliant
w) the device meets the requirements of IEC 61557 standard
x) the product meets EMC requirements (immunity for industrial environment) according to the fol-
lowing standards ............................................................................ EN 61326-1 and EN 61326-2-2
10.2 Additional data
Data on additional uncertainties are useful mainly when the meter is used in non-standard condi-
tions and for metrological laboratories for the purpose of calibration.
10.2.1 Additional uncertainties according to IEC 61557-2 (R
ISO
)
10.2.2 Additional uncertainties according to IEC 61557-3 (Z)
1,2 m lead – 0Ω
5 m lead – 0.011Ω
10 m lead – 0.019Ω
20 m lead – 0.035Ω
WS-03, WS-04 lead – 0.015Ω
Phase angle 0..30°C at the bot-
tom of test range
Network voltage 85%..110%
10.2.3 Additional uncertainties according to IEC 61557-4 (R ±200 mA)